SUTURE ANCHOR VERSUS TOGGLE ROD FOR TOGGLE PIN FIXATION OF COXOFEMORAL LUXATIONS
 |
Coxofemoral luxations belong to the standard problems in small animal practice. Only reposition is not enough in most of the cases - they need a fixation. Which material is used best for the toggle pin fixation, one of the most popular techniques in this problem? A very interesting study from Mississippi State University.
|
|
The mechanical characteristics of toggle rods and Bone BiterTM anchors inserted through the medial acetabular wall for toggle pin repair of coxofemoral luxations were compared in 16 canine cadaver pelves.
No differences were detected in maximum load to failure, displacement at failure, or energy to failure between the two constructs.
Toggle rod constructs failed primarily by breakage of the suture at the rod eyelet.
All of the BoneBiterTM anchor constructs failed when the anchors pulled through the medial acetabular wall.
Source: David Spranklin, Steven Elder, Carolyn Boyle, Ron McLaughlin (2006): Comparison of a Suture Anchor and a Toggle Rod for Use in Toggle Pin Fixation of Coxofemoral Luxations. In: Journal of the American Animal Hospital Association 42:121-126 (2006)
|